Расчеты интраокулярных линз
Рисунок со страницы 46
Рисунок со страницы 46

We had a special Statistics Session. The EC had invited Stan Bentow, PhD (AMO) (below), to give a full 30-min lecture, as a statistician, on how IOL power results should be accurately reported.

Рисунок со страницы 47
Рисунок со страницы 47

The Bootstrap Algorithm

Given a dataset X = (x1, x2, ..., x_n)

  • 1.Select a random sample of size n drawn with replacement from X This is a single bootstrap sample, X*1=x1*1,x2*1,…,0 This B times yielding B bootstrap samples, each of size n

  1. Calculate the parameter of interest for each bootstrap sample. For the mean we would end up with B bootstrap means $\overline{X}^{} = (x_{1}^{}, x_{2}^{}, \ldots, x_{n}^{})$

  • 3.Estimate the mean and standard deviation of the B bootstrap means using the usual equations

  • The bootstrap mean is the original sample mean

  • The bootstrap standard deviation is the standard error of the mean

  1. Use the bootstrap mean and standard error for hypothesis testing, calculating confidence intervals, etc. Note: confidence intervals may be calculated empirically using percentiles (95% confidence interval (2.5 th, 97.5 th))

EDITORIALS

Protocols for Studies of Intraocular Lens Formula Accuracy

Рисунок со страницы 47

KENNETH J. HOFFER, JAIME ARAMBERRI, WOLFGANG HAIGIS, THOMAS OLSEN, GIACOMO SAVINI, H. JOHN SHAMMAS, AND STANLEY BENTOW

M ANY STUDIES HAVE BEEN PUBLISHED ASSESSING the accuracy of intraocular lens (IOL) power calculation. Since the formation of the IOL 'ower Club in 2005, errors have been noted in the protovalues. This may be more difficult with the Haigis formula (see below). It is then appropriate to compare the median absolute error (MedAE) of each formula. The ME being the lowest merely means the lens factor chosen for that He reiterated the opinions of Haigis and Norrby that using the MAE was inappropriate since absolute errors are not a normal Gaussian distribution and that MedAE is more legitimate. However, he detailed methods such as bootstrap­ ping which can be used so that MAE would be legitimate. (This resulted in the IOL Power Club’s accepting an invitation for an Editorial in the American Journal of Ophthalmology in 2015. (Protocols for studies of intraocular lens formula accuracy. Amer J  Ophthalm. 2015;160(3): 403–405)).

First IPC Member Business Meeting

Since the membership had grown over the past decade, the EC decided to have a Membership Business Meeting to allow the members to par­ ticipate and vote on EC decisions. It was the intention of the EC that these official member business meetings continue. During the meeting, Hoffer made a special tribute to Sverker Norrby for his 8 years of service to this Club as a Founder, an EC Member, and a scientific contributor.

2014: Tenth IPC Meeting—Fort Lauderdale, FL, October 8–11, Il Lugano Hotel and Pelican Grand Hotel

The tenth Scientific Meeting was held at the Il Lugano Hotel in Ft. Lauderdale, FL, prior to the AAO Meeting in Chicago.

Attendees: Members: Edmondo Borasio, Jean-Philippe Colliac, Han Bor Fam, Scott McClatchey, and Sabong Srivannaboon. Honorary Member: John Retzlaff. Invited Speaker: Sean Ianchulev, MD.

Рисунок со страницы 48

Industry: Kaspar Baltzer (Haag-Streit), Stan Bentow, PhD (AMO), Tobias Bühren, PhD (Zeiss), Changho Chong, PhD (MOVU), Alessandro Foggi and Pierre Gelissen (Topcon EU), Martin Gründig, PhD (Alcon), Hiryuki Hiramatsu (Nidek), Jörg Iwanczuk (Oculus), Chihiro Kato, Koki Nishiwaki, and Hisashi Onizaki (all Tomey), Thomas Padrick, PhD (WaveTec), Gregor Schmid, PhD (Ziemer), and Francesco Versaci (CSO).

Рисунок со страницы 48