Расчеты интраокулярных линз
Рисунок со страницы 46
Рисунок со страницы 46

У нас была специальная «Сессия по статистике». EC пригласил Stan Bentow, PhD (AMO) (ниже), чтобы он, как статистик, прочитал полноценную 30-минутную лекцию о том, как следует точно представлять результаты по силе ИОЛ.

We had a special Statistics Session. The EC had invited Stan Bentow, PhD (AMO) (below), to give a full 30-min lecture, as a statistician, on how IOL power results should be accurately reported.

Рисунок со страницы 47
Рисунок со страницы 47

Бутстреп-алгоритм

The Bootstrap Algorithm

Дан набор данных X = (x1, x2, ..., x_n)

Given a dataset X = (x1, x2, ..., x_n)

  • 1. Выберите случайную выборку размера n, извлеченную из X с возвращением. Это одна бутстреп-выборка, X*1=x1*1,x2*1,…,0 Это повторяется B раз, что дает B бутстреп-выборок, каждая размером n

    1.Select a random sample of size n drawn with replacement from X This is a single bootstrap sample, X*1=x1*1,x2*1,…,0 This B times yielding B bootstrap samples, each of size n

  1. Рассчитайте интересующий параметр для каждой бутстреп-выборки. Для среднего значения мы получим B бутстреп-средних $\overline{X}^{} = (x_{1}^{}, x_{2}^{}, \ldots, x_{n}^{})$

    Calculate the parameter of interest for each bootstrap sample. For the mean we would end up with B bootstrap means $\overline{X}^{} = (x_{1}^{}, x_{2}^{}, \ldots, x_{n}^{})$

  • 3. Оцените среднее значение и стандартное отклонение для B бутстреп-средних, используя обычные уравнения

    3.Estimate the mean and standard deviation of the B bootstrap means using the usual equations

  • Бутстреп-среднее соответствует среднему значению исходной выборки

    The bootstrap mean is the original sample mean

  • Бутстреп-стандартное отклонение представляет собой стандартную ошибку среднего

    The bootstrap standard deviation is the standard error of the mean

  1. Используйте бутстреп-среднее и стандартную ошибку для проверки гипотез, расчета доверительных интервалов и т. д. Примечание: доверительные интервалы могут быть рассчитаны эмпирически с использованием процентилей (95% доверительный интервал (2.5 -й, 97.5 -й))

    Use the bootstrap mean and standard error for hypothesis testing, calculating confidence intervals, etc. Note: confidence intervals may be calculated empirically using percentiles (95% confidence interval (2.5 th, 97.5 th))

РЕДАКЦИОННЫЕ СТАТЬИ

EDITORIALS

Протоколы исследований точности формул расчета интраокулярных линз

Protocols for Studies of Intraocular Lens Formula Accuracy

Рисунок со страницы 47

КЕННЕТ J. ХОФФЕР, ХАЙМЕ АРАМБЕРРИ, ВОЛЬФГАНГ HAIGIS, ТОМАС ОЛСЕН, ДЖАКОМО САВИНИ, H. ДЖОН ШАММАС И СТЭНЛИ БЕНТОУ

KENNETH J. HOFFER, JAIME ARAMBERRI, WOLFGANG HAIGIS, THOMAS OLSEN, GIACOMO SAVINI, H. JOHN SHAMMAS, AND STANLEY BENTOW

ОПУБЛИКОВАНО МНОГО ИССЛЕДОВАНИЙ, оценивающих точность расчета силы интраокулярных линз (ИОЛ). С момента основания ИОЛ (интраокулярная линза) Power Club в 2005 году в протоколах были отмечены ошибки в значениях. Это может быть более затруднительно с формулой Haigis (см. ниже). В таком случае целесообразно сравнить медианную абсолютную ошибку (MedAE) каждой формулы. Тот факт, что ME является самой низкой, означает лишь, что выбранный для этого коэффициент линзы был наиболее подходящим. Было подтверждено мнение Haigis и Norrby о том, что использование MAE было неуместным, поскольку абсолютные ошибки не имеют нормального распределения Гаусса и что MedAE является более обоснованным показателем. Однако он подробно описал такие методы, как бутстрэппинг, которые могут быть использованы для того, чтобы MAE стал обоснованным. (Это привело к тому, что ИОЛ, интраокулярная линза Power Club принял приглашение написать редакционную статью в American Journal of Ophthalmology в 2015 г. (Protocols for studies of intraocular lens formula accuracy. Amer J Ophthalm. 2015;160(3): 403–405)).

M ANY STUDIES HAVE BEEN PUBLISHED ASSESSING the accuracy of intraocular lens (IOL) power calculation. Since the formation of the IOL 'ower Club in 2005, errors have been noted in the protovalues. This may be more difficult with the Haigis formula (see below). It is then appropriate to compare the median absolute error (MedAE) of each formula. The ME being the lowest merely means the lens factor chosen for that He reiterated the opinions of Haigis and Norrby that using the MAE was inappropriate since absolute errors are not a normal Gaussian distribution and that MedAE is more legitimate. However, he detailed methods such as bootstrap­ ping which can be used so that MAE would be legitimate. (This resulted in the IOL Power Club’s accepting an invitation for an Editorial in the American Journal of Ophthalmology in 2015. (Protocols for studies of intraocular lens formula accuracy. Amer J  Ophthalm. 2015;160(3): 403–405)).

Первое деловое собрание членов IPC

First IPC Member Business Meeting

Поскольку за последнее десятилетие число членов увеличилось, EC решил провести деловое собрание членов, чтобы позволить им участвовать и голосовать по решениям EC. Намерением EC было продолжение этих официальных деловых собраний членов. В ходе встречи Hoffer выразил особую признательность Sverker Norrby за 8 лет службы в этом Клубе в качестве основателя, члена EC и научного сотрудника.

Since the membership had grown over the past decade, the EC decided to have a Membership Business Meeting to allow the members to par­ ticipate and vote on EC decisions. It was the intention of the EC that these official member business meetings continue. During the meeting, Hoffer made a special tribute to Sverker Norrby for his 8 years of service to this Club as a Founder, an EC Member, and a scientific contributor.

2014: Десятая встреча IPC — Форт-Лодердейл, Флорида, 8–11 октября, отели Il Lugano и Pelican Grand

2014: Tenth IPC Meeting—Fort Lauderdale, FL, October 8–11, Il Lugano Hotel and Pelican Grand Hotel

Десятая научная встреча была проведена в отеле Il Lugano в Форт-Лодердейле, Флорида, перед встречей AAO в Чикаго.

The tenth Scientific Meeting was held at the Il Lugano Hotel in Ft. Lauderdale, FL, prior to the AAO Meeting in Chicago.

Участники: Члены: Edmondo Borasio, Jean-Philippe Colliac, Han Bor Fam, Scott McClatchey и Sabong Srivannaboon. Почетный член: John Retzlaff. Приглашенный докладчик: Sean Ianchulev, MD.

Attendees: Members: Edmondo Borasio, Jean-Philippe Colliac, Han Bor Fam, Scott McClatchey, and Sabong Srivannaboon. Honorary Member: John Retzlaff. Invited Speaker: Sean Ianchulev, MD.

Представители индустрии: Kaspar Baltzer (Haag-Streit), Stan Bentow, PhD (AMO), Tobias Bühren, PhD (Zeiss), Changho Chong, PhD (MOVU), Alessandro Foggi и Pierre Gelissen (Topcon EU), Martin Gründig, PhD (Alcon), Hiryuki Hiramatsu (Nidek), Jörg Iwanczuk (Oculus), Chihiro Kato, Koki Nishiwaki и Hisashi Onizaki (все из Tomey), Thomas Padrick, PhD (WaveTec), Gregor Schmid, PhD (Ziemer) и Francesco Versaci (CSO).

Industry: Kaspar Baltzer (Haag-Streit), Stan Bentow, PhD (AMO), Tobias Bühren, PhD (Zeiss), Changho Chong, PhD (MOVU), Alessandro Foggi and Pierre Gelissen (Topcon EU), Martin Gründig, PhD (Alcon), Hiryuki Hiramatsu (Nidek), Jörg Iwanczuk (Oculus), Chihiro Kato, Koki Nishiwaki, and Hisashi Onizaki (all Tomey), Thomas Padrick, PhD (WaveTec), Gregor Schmid, PhD (Ziemer), and Francesco Versaci (CSO).

Рисунок со страницы 48